Simple and reliable analyzer for continuous dust measurement.
Simple and reliable analyzer for continuous dust measurement. The narrow laser beam transmission method minimizes the effects of scattering and reflection. Unique optical channel – “zero tube”, for continuous zero update and calibration several times within a minute. Continuous monitoring of optical surfaces and contamination compensation. Unique calibration system with simulated 0 and span – adjustable period between calibrations. Easy installation and reduced operating costs. – Calibration is carried out by a “reference filter”, does not require additional equipment and instruments. – Calibration does not require removal of the device from the pipe – No gravimetric comparative measurement required – Calibration is carried out together with the controller – monitor. – Calibration filters have calibration data. – The results of the last 3 calibrations are saved and can be printed.