328А Steam Sterilized рН Sensor

328А Steam Sterilized рН Sensor
Manufacturer:
Emerson

328А model pH sensor produced by Rosemount Analytical is designed for рН measurements where sensors shall be steam-sterilized at a temperature up to 130°C. 328А sensor has a ceramic body with silicon rubber insulation providing perfect chemical resistance to substances applied in food and pharmaceutical industries. Sensor can resist up to 50 steam sterilization cycles. 328А model has dual reference cell protecting reference electrode against toxic effect and pollution.

More
Category: Steam-sterilized sensors

328А model pH sensor produced by Rosemount Analytical is designed for рН measurements where sensors shall be steam-sterilized at a temperature up to 130°C. 328А sensor has a ceramic body with silicon rubber insulation providing perfect chemical resistance to substances applied in food and pharmaceutical industries. Sensor can resist up to 50 steam sterilization cycles. 328А model has dual reference cell protecting reference electrode against toxic effect and pollution.

Technological information and specifications
Transmitter Type Flow-type
Wetted materials Ceramics, silicone, glass
Automatic temperature compensation Ordered separately (manufacturer advice required)
Temperature / nominal pressure 446 kPa (absolute) at 130°C

584 kPa (absolute) at 80°C

791 kPa (absolute) at 40°C

Cable Two-core low-noise coaxial cable
Process connection Standard 12mm cut-in fitting
Operating performance Sensor can resist up to 50 steam sterilization cycles.
Accuracy ± 0.02pH
Measurement range 0-13 pH 1.67-4.01 pH  96% accuracy

4.01-12.0 pH  99% accuracy

12.0-13.0 pH  97% accuracy

Convergence ± 0.05pH

328А model pH sensor produced by Rosemount Analytical is designed for рН measurements where sensors shall be steam-sterilized at a temperature up to 130°C. 328А sensor has a ceramic body with silicon rubber insulation providing perfect chemical resistance to substances applied in food and pharmaceutical industries. Sensor can resist up to 50 steam sterilization cycles. 328А model has dual reference cell protecting reference electrode against toxic effect and pollution.

Technological information and specifications
Transmitter Type Flow-type
Wetted materials Ceramics, silicone, glass
Automatic temperature compensation Ordered separately (manufacturer advice required)
Temperature / nominal pressure 446 kPa (absolute) at 130°C

584 kPa (absolute) at 80°C

791 kPa (absolute) at 40°C

Cable Two-core low-noise coaxial cable
Process connection Standard 12mm cut-in fitting
Operating performance Sensor can resist up to 50 steam sterilization cycles.
Accuracy ± 0.02pH
Measurement range 0-13 pH 1.67-4.01 pH  96% accuracy

4.01-12.0 pH  99% accuracy

12.0-13.0 pH  97% accuracy

Convergence ± 0.05pH
Our contacts
pin_drop
г. Москва, Бережковская набережная,16А стр.3
schedule
пн-пт с 9 до 18
Отдел сервисного обслуживания
pin_drop
г. Дубна, ул. Университетская, д. 11, стр. 14
schedule
пн-пт с 9 до 18
Отдел сервисного обслуживания

г. Москва info@ecohimpribor.ru

г. Санкт-Петербург spb@ecohimpribor.ru

г. Краснодар krd@ecohimpribor.ru

г. Казань kzn@ecohimpribor.ru

г. Омск oms@ecohimpribor.ru

г. Новосибирск nsk@ecohimpribor.ru

г. Иркутск irk@ecohimpribor.ru

г. Пермь prm@ecohimpribor.ru

    Contact us