1057 transmitter has three inputs from sensors and four current outputs that allows to reduce consumption by one circuit and saves panel space.
Input board for pH signal supports pH, ORP measurements and measurements using ion selective electrodes (ISE).
Input board for conductivity signal supports measurement of conductivity and specific resistance, determination of total content of dissolved solids, salinity, as well as allows to plot a graph of percentage concentration for specific applications.
Thanks to modular structure, signal input boards can be replaced in field conditions that simplifies configuration changes. For convenience purposes, real-time measurement data are always shown on the display during adjustment and calibration.
Standard structural features includes insulated inputs, 7 in-built languages, four current outputs 4-20 mA, 4 signaling relays.
Analyzer has in-built preamplifier with SMART technology for electrochemical analog sensors.
1057 transmitter has three inputs from sensors and four current outputs that allows to reduce consumption by one circuit and saves panel space.
Input board for pH signal supports pH, ORP measurements and measurements using ion selective electrodes (ISE).
Input board for conductivity signal supports measurement of conductivity and specific resistance, determination of total content of dissolved solids, salinity, as well as allows to plot a graph of percentage concentration for specific applications.
Thanks to modular structure, signal input boards can be replaced in field conditions that simplifies configuration changes. For convenience purposes, real-time measurement data are always shown on the display during adjustment and calibration.
Standard structural features includes insulated inputs, 7 in-built languages, four current outputs 4-20 mA, 4 signaling relays.
Analyzer has in-built preamplifier with SMART technology for electrochemical analog sensors.
Technological information and specifications of device |
|||
Casing material |
Polycarbonate. NEMA 4X/CSA 4 |
||
Dust and moisture protection |
IP65 |
||
Cable conduit openings |
It is possible to use ½ inch fitting or PG13.5. |
||
Display |
Monochrome graphic LCD. Resolution: 128 x 96 pxl. Display backlight. Active display surface: 58 x 78 mm (2.3 x 3.0 inch). |
||
Ambient temperature and humidity |
From 0 to 55 C, relative humidity from 5 to 95 % (without condensation). |
||
Storage temperature |
from -20 to 60°С |
||
Power supply |
Optional |
02 |
24 VDC and 4 relays |
03 |
Switch-selectable AC 85-265VAC, 50/60 Hz, 4 signaling relays |
||
Relay |
With break-make contact (C type), single-pole double-throw (SPDT), impeded into epoxied resin. |
||
Input |
Up to three electrically insulated inputs of signals from sensors. |
||
Output |
Four electrically isolated current outputs 4-20 mA or 0-20 mA. Extensibility Maximum load: 550 Ohm |
||
Accuracy of current output |
± 0.05 mA at 25 °C |
||
Camp connection specifications |
Power socket (3 outputs): 24-12 AWG wires. Connector blocks on signal board: 26-16 AWG wires. Power socket connectors (4 outputs): 24-16 AWG wires. Connector blocks on signal relay: 24-12 AWG wires. |
||
Weight/shipping weight |
(rounded to 0.5 kg): 1.5 /2.0 kg |
||
Dimension |
155 x 155 x 131 mm |
г. Москва info@ecohimpribor.ru
г. Санкт-Петербург spb@ecohimpribor.ru
г. Краснодар krd@ecohimpribor.ru
г. Казань kzn@ecohimpribor.ru
г. Омск oms@ecohimpribor.ru
г. Новосибирск nsk@ecohimpribor.ru
г. Иркутск irk@ecohimpribor.ru
г. Пермь prm@ecohimpribor.ru